Cena s DPH / bez DPH
Hlavní stránka>BS EN 10315:2006 Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique
Sponsored link
sklademVydáno: 2009-09-30
BS EN 10315:2006 Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique

BS EN 10315:2006

Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique

Formát
Dostupnost
Cena a měna
Anglicky Zabezpečené PDF
K okamžitému stažení
6448 Kč
Čtěte normu po dobu 1 hodiny. Více informací v kategorii E-READING
Čtení normy
na 1 hodinu
644.80 Kč
Čtěte normu po dobu 24 hodin. Více informací v kategorii E-READING
Čtení normy
na 24 hodin
1934.40 Kč
Anglicky Tisk
Skladem
6448 Kč
Označení normy:BS EN 10315:2006
Počet stran:32
Vydáno:2009-09-30
ISBN:978 0 580 68841 6
Status:Standard
Popis

BS EN 10315:2006


This standard BS EN 10315:2006 Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique is classified in these ICS categories:
  • 77.140.20 Stainless steels
  • 77.040.30 Chemical analysis of metals

This European Standard specifies a procedure on how to improve the performance of a routine XRF method, already in use for analysis of high alloy steels, by using a ”near by technique”.

The ”near by technique” requires at least one target sample (preferable a CRM) of a similar composition as the unknown sample.

The method is applicable to elements within the concentration ranges according to Table 1:

Table 1 — Concentration ranges

Element Concentration range, % (m/m) a
Si 0,05 to 1,5
Mn 0,05 to 5,0
P 0,005 to 0,035
Cr 10 to 25
Ni 0,1 to 30
Mo 0,1 to 6,5
Cu 0,02 to 1,5
Co 0,015 to 0,30
V 0,015 to 0,15
Ti 0,015 to 0,50
Nb 0,05 to 1,0

a The concentration ranges specified, represents those ranges studied during the precision test. The procedure has the potential to be used outside those ranges but it needs to be validated by each laboratory in every case.

The method is applicable to analysis of either chill-cast or wrought samples having a diameter of at least 25 mm and with a carbon concentration of less than 0,3 % (see NOTE). Other elements should have a concentration below 0,2 %.

NOTE High carbon concentrations, in combination with high Mo and Cr concentrations, could have undesirable structural effects on the sample and could affect the determination of phosphorus and chromium, in particular.

Matrix effects exist between the elements listed. To compensate for those inter-element effects, mathematical corrections shall be applied. A variety of computer programs for corrections is commonly used and included in the software package from the manufacturers.