Menu
0
Total price
0 €
PRICES include / exclude VAT
Homepage>PD CEN/TS 18073:2024 Ambient air. Determination of lung deposited surface area (LDSA) concentration using aerosol monitors based on diffusion charging
Sponsored link
sklademVydáno: 2024-09-27
PD CEN/TS 18073:2024 Ambient air. Determination of lung deposited surface area (LDSA) concentration using aerosol monitors based on diffusion charging

PD CEN/TS 18073:2024

Ambient air. Determination of lung deposited surface area (LDSA) concentration using aerosol monitors based on diffusion charging

Format
Availability
Price and currency
Anglicky Secure PDF
Immediate download
298.32 €
You can read the standard for 1 hour. More information in the category: E-reading
Reading the standard
for 1 hour
29.83 €
You can read the standard for 24 hours. More information in the category: E-reading
Reading the standard
for 24 hours
89.50 €
Anglicky Hardcopy
In stock
298.32 €
Označení normy:PD CEN/TS 18073:2024
Počet stran:36
Vydáno:2024-09-27
ISBN:978 0 539 25287 3
Status:Standard
DESCRIPTION

PD CEN/TS 18073:2024


This standard PD CEN/TS 18073:2024 Ambient air. Determination of lung deposited surface area (LDSA) concentration using aerosol monitors based on diffusion charging is classified in these ICS categories:
  • 13.040.20 Ambient atmospheres
This document specifies a process for the electrical diffusion charging of aerosols with subsequent measurement of particle charge. With the aid of this method, it is possible to determine the lung-deposited surface area (LDSA) concentration of particles in ambient air. Depending on the design of the electrical diffusion charger, the LDSA of particles in the size range of approximately 20 nm to approximately 300 nm is measurable. Furthermore, this document specifies design criteria for LDSA measuring aerosol monitors as well as performance criteria and the associated test procedures. The performance criteria depend on the application and they are more stringent when the instrument is operated in an air quality monitoring station. In the determination of the LDSA concentration, the share of geometric particle surface area concentration is determined that can be deposited in the alveolar region of the human lung. Typical particle surface area concentrations with alveolar deposition measured in urban areas range from 5 µm2/cm3 to 50 µm2/cm3. Instruments based on this measurement principle can be designed to be very compact with a low power consumption. This makes them ideally suited for handheld measurements, other forms of mobile application or to measure personal exposure. On the other hand, they can be easily adapted to serve as a stationary instrument in air quality monitoring stations.